Leakage current and catheterization protection

Short answer

Leakage current is the unwanted small current flowing through insulation and stray capacitance to the chassis or ground. For a catheterized heart patient, protect by low-resistance grounding, equipotential bonding, and isolation so leakage bypasses the patient. With 100 µA leakage, 0.1 Ω ground wire, and 1 kΩ patient, patient current is 10 nA, safe.

What is leakage currents? How a Heart patient who is under catheterization procedure is protected from the leakage current? Assume the values of leakage current and patient resistance with suitable ground wire for proper grounding.

Leakage current is the unwanted small current that flows from the live parts of an electrical device through its insulation and stray capacitance to the metal chassis (case), and then to the ground, or through a person touching it. Even with no fault, 50/60 Hz mains produces tens of µA through stray capacitance between the power wires and the chassis. Types: enclosure (chassis) leakage, patient leakage, and patient auxiliary leakage.

Limits: for general patient equipment, 100 µA (chassis); for devices with direct cardiac connection (catheter, pacing wire), the limit is 10 µA, because microshock of about 100 µA through the heart can cause ventricular fibrillation.

Protection of a catheterised heart patient

  1. Grounding of the chassis through a low-resistance ground wire (three-wire system) so the leakage current flows to the ground and not through the patient.
  2. Equipotential grounding: all metal surfaces near the patient (bed, instruments, pipes) are connected by heavy wires to a common ground point, so there is no potential difference between them.
  3. Isolation transformer / isolated power (with a line isolation monitor), so the first fault does not cause shock.
  4. Isolated (floating) patient connections: isolation amplifiers (transformer or optical) with very small coupling capacitance.
  5. Double insulation and battery-operated equipment; GFCI/earth leakage breaker; regular testing of leakage.
  6. Catheter ends insulated, staff handling with dry gloves, and only one device connected to the catheter at a time.

Example calculation (assumed values)

 Assume: chassis leakage current  Ileak = 100 uA
         ground wire resistance   Rg    = 0.1 ohm
         patient (catheter-to-ground body path) Rp = 1 kohm

With the ground wire intact, the chassis voltage is Vc = Ileak × Rg = 100 µA × 0.1 Ω = 10 µV. The current through the patient is Ip = Vc/Rp = 10 µV/1 kΩ = 10 nA, far below the 10 µA limit; so almost all the leakage current goes to the ground wire.

If the ground wire breaks (Rg → ∞), the whole 100 µA can flow through the patient: Ip ≈ 100 µA, which is 10 times the safe limit and may produce fibrillation.

Hence a good ground wire (Rg < 0.1 Ω, thick copper), equipotential grounding and periodic checks give the necessary protection.