Data logger input scanner and conditioner design
Short answer
For the scanner, use a multiplexer with per-channel sampling rate of 40 kHz (scan rate 120 kHz) and anti-alias filters at 16 kHz. The signal conditioner must amplify by 50 and add a 2.5 V offset to map ±50 mV to 0–5 V, using an instrumentation amplifier and a summing stage.
Signals from three different transducers need to be recorded in a data logger. The analog signals supplied by the three transducers are dual polarity (-50 mV to 50 mV) having frequencies of 5 KHz, 10 KHz and 15 KHz. Explain the design of the following stages of the data logger: (a) Input scanner stage of the data logger such that it can appropriately sample the incoming signals. (b) Signal conditioner stage if the 8-bit ADC used inside the data logger accepts only positive polarity signals ranging from 0 volts to 5 volts.
Given
Three transducer outputs, each +/-50 mV (100 mV span), with frequencies 5, 10 and 15 kHz. ADC: 8-bit, accepts 0 to 5 V.
(a) Input scanner stage
The input scanner is an analog multiplexer (CD4051/DG408, 8 channels) controlled by the logger's controller; it connects each transducer to the S/H and ADC in turn.
- Sampling rate per channel: by the Nyquist theorem, fs >= 2 x fmax = 2 x 15 kHz = 30 kHz. Choose 40 kHz per channel (a margin for the anti-alias filter roll-off) for all three channels so that each is sampled correctly.
- Scanning rate: 3 channels x 40 kHz = 120 kHz, so each channel slot is 1 / 120 kHz = 8.33 us. The S/H acquisition time + ADC conversion time must be less than 8.33 us (fast SAR or flash ADC such as ADC0820, 1.5 us conversion; the S/H must have an acquisition time below 3 us).
- Use an anti-alias low-pass filter (cut-off about 16 kHz, 4th order) at each channel input, before the multiplexer.
- Use a sample-and-hold (one common after the MUX, or one per channel for simultaneous sampling) with aperture time much smaller than the period; for 15 kHz and an amplified 2.5 V peak and 8 bits: Dt <= (LSB/2)/(2 pi f A) = 9.8 mV / (2 pi x 15e3 x 2.5) = 41 ns.
- Channel selection by counter or controller: address lines A0, A1 of the MUX step through 0, 1, 2 repeatedly.
T1 -> [LPF 16k] -+
T2 -> [LPF 16k] -+-> [4051 MUX] -> [S/H] -> to conditioner/ADC
T3 -> [LPF 16k] -+ ^ select (counter, 120 kHz)
(b) Signal conditioner stage
The ADC accepts only 0 to 5 V, but the signal is -50 mV to +50 mV. Both amplification and level shifting are needed.
- Gain = output span / input span = 5 V / 0.1 V = 50.
- Offset: -50 mV must give 0 V, +50 mV must give 5 V, so the mid-point 0 V must give 2.5 V. Therefore Vout = 50 x Vin + 2.5 V.
- Check: Vin = -50 mV -> 0 V; Vin = 0 -> 2.5 V; Vin = +50 mV -> 5 V.
Vin -> [Instrumentation amp, gain 50] -> [Summing amp: +2.5 V] -> ADC
(INA114, Rg = 1.02 kohm) (2.5 V from a reference IC)
Implementation: an instrumentation amplifier INA114 with Rg = 50 kohm/(G - 1) = 1.02 kohm gives gain 50 with good common-mode rejection. The second stage is a summing amplifier (or a differential amplifier with reference) which adds a dc level of +2.5 V from a precision reference (REF02 + divider). Clamp diodes at the ADC input and a final low-pass filter protect the ADC.
✓ Answer checked by Claude Sonnet 5.5
Related questions
- Data logger with Bluetooth scatternet · Instrumentation II, chapter 5
- Three transducers to control room: routing and S/H · Instrumentation II, chapter 4
- Why digitize analog signals · Instrumentation II, chapter 4 · asked 2×
- Data logger block diagram and operation · Instrumentation II, chapter 5 · asked 6×
- Characteristics of a data logger · Instrumentation II, chapter 5 · asked 6×
- Compact data logger characteristics and block diagram · Instrumentation II, chapter 5
- Analog communication block diagram; FDM of three signals · Communication Systems, chapter 5
- TST vs STS and three-stage design · Telecommunications, chapter 4
- Design 200×200 three-stage switch · Telecommunications, chapter 4
- Dual slope ADC numerical (12 kHz clock) · Advanced Electronics, chapter 3