- 2075 Chaitra · 8 marks
Dialysis fluid mixing and electro-hydraulic ESWL
Short answer
Dialysate: RO water and concentrate mixed (34:1), warmed, degassed, then passed counter-current through dialyzer for diffusion/ultrafiltration. ESWL: A high-voltage capacitor discharges across a spark-gap at the first focus of an ellipsoidal reflector, generating focused shockwaves that shatter kidney stones at the second focus, guided by X-ray/ultrasound.
How dialysate fluid is used with RO water in mixing chamber to artificial kidney during dialysis? Draw a lay-out diagram of Electro-Hydraulic based ESWL with its required major components.
Dialysate preparation with RO water
In haemodialysis the artificial kidney (dialyser) removes waste (urea, creatinine) and excess water from the blood across a semi-permeable membrane into the dialysate.
Tap water -> [Pre-treatment: filter, softener, carbon]
-> [Reverse osmosis (RO) unit] -> RO water
Dialysate concentrate --[Proportioning pump]--+
RO water --------------[Proportioning pump]--+-> [MIXING CHAMBER]
|
[Heater 37 C]<-[Deaerator]<-[Conductivity/pH monitor]
|
[Flow meter] -> DIALYSER <-> blood (counter-current)
|
[Blood-leak detector] -> drain
Working:
- RO water, free of ions, bacteria and chemicals, enters the mixing chamber.
- The concentrate (sodium, potassium, calcium and magnesium chlorides, bicarbonate, glucose) is added by a proportioning pump, normally 1 part concentrate to 34 parts water.
- The mixture is warmed to about 37 °C and degassed (deaerator) so that bubbles do not form in the dialyser.
- Conductivity and pH sensors check that the mixture is correct; if not, the fluid is sent to the drain (bypass).
- The dialysate flows (about 500 mL/min) through the dialyser in the direction opposite to the blood, so waste diffuses from the blood into the dialysate, and the pressure difference removes extra water (ultrafiltration). A blood-leak detector checks the used fluid for blood.
Electro-hydraulic ESWL (extracorporeal shock wave lithotripsy)
ESWL breaks kidney stones without surgery using focused shock waves.
[HV supply 10-20 kV]--[Capacitor]--[Switch/trigger]
|
[Spark-gap electrode F1]
water-filled ellipsoidal reflector
/ shock waves \
/ \
v focus F2 v
Patient (in water bath/cushion) -> kidney STONE at F2
[X-ray (fluoroscopy)/ultrasound localisation] -> [Control unit]
Components: HV generator and capacitor, spark gap (underwater electrode) at the first focus F1 of a half-ellipsoid reflector, water medium, coupling cushion, an imaging system (X-ray or ultrasound) to position the stone at the second focus F2, and the control unit with ECG triggering. Working: the capacitor discharges across the electrodes, which vaporises water and creates a shock wave. The reflector focuses the shock waves on the stone at F2. Repeated shocks (1000 to 3000) crush the stone into small fragments, which pass out in the urine.
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Man-instrument system and ECG telemetry layout · Biomedical Instrumentation, chapter 1
- Is a hydraulic jump stilling basin required? · Hydropower Engineering, chapter 4
- Entropy change on mixing ice and water · Fundamental of Thermodynamics & Heat Transfer, chapter 5
- Mixing 12 kg water and 8 kg ice: entropy change · Fundamental of Thermodynamics & Heat Transfer, chapter 5
- Major components of a power station · Power System Analysis I, chapter 1 · asked 2×