Short answer

The 12-lead ECG uses 10 electrodes: RA, LA, LL, RL (ground), and V1-V6 chest positions. It records 3 bipolar limb leads (I, II, III), 3 augmented unipolar leads (aVR, aVL, aVF), and 6 precordial leads (V1-V6) referenced to Wilson's central terminal. Electrode types include limb plate electrodes, suction chest electrodes, disposable Ag/AgCl adhesive electrodes, and floating electrodes to reduce motion artifacts. Skin preparation with conductive paste lowers contact impedance.

Draw a set of 12 leads ECG recording configuration with proper connection diagram. What types of electrodes are used to record ECG?

The standard 12-lead ECG consists of 3 bipolar limb leads, 3 augmented unipolar limb leads and 6 unipolar chest (precordial) leads.

Electrode placement

  • RA: right arm, LA: left arm, LL: left leg, RL: right leg (ground, reference).
  • Chest V1 to V6: V1 – 4th intercostal space right of sternum; V2 – 4th space left of sternum; V3 – midway between V2 and V4; V4 – 5th space at the mid-clavicular line; V5 – same level as V4 on the anterior axillary line; V6 – same level at the mid-axillary line.

Leads

GroupLeadsMeasures
Bipolar limbI (LA−RA), II (LL−RA), III (LL−LA)Einthoven triangle
Augmented limbaVR, aVL, aVFone limb vs average of other two
ChestV1–V6one chest electrode vs Wilson's central terminal

Connection diagram

  RA --+
  LA --+--> [Lead selector] --> [Differential] --> [Filter] --> [Driver] --> Recorder
  LL --+       switch            amplifier
  V1..V6 ----^
  RL ---------> ground (right-leg drive)

  Wilson central terminal:
     RA --[5k]--+
     LA --[5k]--+--> CT (V = (RA+LA+LL)/3) --> reference for V leads
     LL --[5k]--+

Wilson's central terminal averages the three limb potentials and serves as the indifferent reference for the chest leads. For aVR, aVL and aVF, the limb electrode is the active one and the other two are joined through resistors (Goldberger).

Electrodes used for ECG

  1. Limb plate electrodes: metal plates (German silver) strapped on the arms and legs with electrolyte jelly.
  2. Suction (chest) electrodes: a metal cup with a rubber bulb, used for the precordial leads.
  3. Disposable adhesive Ag/AgCl electrodes: foam pads with gel, used for monitoring and stress testing.
  4. Floating electrodes: a metal disk recessed in a cup so that the metal does not touch the skin directly (contact is through electrolyte paste), which reduces motion artefacts.
  5. Dry/conducting-rubber and flexible electrodes for long-term use.

Skin preparation and electrode jelly (conductive paste) are used to reduce the contact impedance.