Piezoelectric pulse transmitter, A-mode vs B-mode

Short answer

A high-voltage supply charges a capacitor; a fast switch discharges it through the crystal, producing a brief ultrasound pulse. Echoes from tissue boundaries are received; depth is computed by d = c·t/2. A-mode shows signal amplitude vs depth as spikes, while B-mode displays brightness-modulated dots to form a cross-sectional image.

How a transmitter circuit discharges a capacitor through the crystal of a piezo-electric transducer to generate large amplitude of pulses? How these pulses are utilized to view internal structures of the human body? Compare A-mode and b-mode display format.

Transmitter circuit

The piezoelectric crystal needs a short, high-amplitude electrical pulse to produce a strong ultrasound pulse.

 HV supply --[R]--+--[C]--+-------------+--> T/R switch --> receiver amp
                  |       |             |
           [SCR / switch] (crystal)  [R damp]
                  |       |             |
                 gnd     gnd           gnd
                  ^
           trigger pulse (clock)

Working: the capacitor C charges slowly through the resistor R from a high-voltage supply (hundreds of volts). A trigger from the clock turns on the fast switch (SCR, thyratron or MOSFET), so C discharges almost instantly through the crystal (and a damping resistor). The sharp high-voltage spike shocks the crystal into oscillation at its resonant frequency (1 to 10 MHz), producing a short damped burst of ultrasound (about 1 µs). The cycle repeats at the pulse repetition frequency (1 to 5 kHz). Between pulses, the same crystal acts as the receiver for the echoes.

Use of pulses to view internal structures (pulse-echo)

  1. The transducer, placed on the skin with coupling gel, sends a pulse into the body.
  2. At each boundary between tissues of different acoustic impedance, part of the pulse is reflected as an echo.
  3. The echo returns to the crystal, which converts it to an electrical signal; this is amplified (with time-gain compensation) and detected.
  4. The depth of the reflector is d = c·t/2 (c = 1540 m/s, t = time between pulse and echo); the echo strength shows the reflectivity.
  5. The signals are displayed in A, B or M mode to show the organs (liver, kidney, foetus, heart).

A-mode versus B-mode

PointA-mode (amplitude)B-mode (brightness)
DisplaySpikes: amplitude vs depthDots of brightness: grey-scale image
DimensionOne-dimensionalTwo-dimensional cross-section
TransducerFixed, singleMoved or scanned (or array)
Echo strength shown byHeight of the peakBrightness of the dot
ImageNo pictureAnatomical picture
UsesDistance measurement, eye (ophthalmology), brain midlineObstetrics, abdomen, heart, tissue imaging
ComplexitySimple, cheapMore complex (scan converter and memory)