- Practice · 6 marks
Optical lever comparator magnification
Short answer
Magnification M = 2L/x = (2×250)/5 = 100. Least count = 0.5/100 = 0.005 mm. Deviation = 3.2×0.005 = 0.016 mm, so shaft size = 25.016 mm; it exceeds the upper limit 25.015 mm and is rejected (oversize by 0.001 mm).
A mechanical-optical comparator uses a mirror pivoted at one end and pushed by the plunger at a distance of 5 mm from the pivot. The reflected beam falls on a scale 250 mm from the mirror. Derive the magnification of this single optical lever and find it. If the scale divisions are 0.5 mm apart, what is the least count? The instrument is set to zero with a 25.000 mm slip gauge. A shaft is then put in and the reading is +3.2 divisions. Find the size of the shaft, and say whether it is within the limits mm.
Derivation
Let the plunger move by and the mirror rotate by about the pivot, with the plunger acting at distance from the pivot:
A reflected ray turns through when the mirror turns by . On a scale at distance the spot moves by
Least count
Shaft size
The deviation from the master is mm, so
Limits are to mm. The shaft is larger than the upper limit by mm, therefore it is not acceptable (oversize, can be reworked).
Answer: , LC = 0.005 mm, size = 25.016 mm, rejected (oversize by 0.001 mm).
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Classification and merits of comparators · Metrology, chapter 7
- Vernier caliper and micrometer readings · Metrology, chapter 4
- Straightness of a bed using a spirit level · Metrology, chapter 13
- Fringe measurement of flatness error · Metrology, chapter 8
- Spirit level sensitivity and tilt · Metrology, chapter 5
- Sine bar principle, setting and error · Metrology, chapter 5
- Rod pivoted at O, spring compressed 25 mm · Applied Mechanics (Dynamics), chapter 6
- Slab three edges continuous, one short edge discontinuous (5x4 m, 250 beam) · Design of RCC Structure, chapter 9 · asked 3×
- Crank and slotted lever velocity analysis · Theory of Mechanism and Machine I, chapter 7
- Load sharing: 250/500 MW, frequency falls to 49.5 Hz · Power Plant Equipment, chapter 2 · asked 2×