- Practice · 8 marks
Calibration line and nonlinearity of transducer
Short answer
Fit y = mx + c by least squares; sensitivity (slope) = 0.499 V/bar, zero offset = 0.0138 V. Maximum deviation is 0.0419 V at 4 bar, full-scale output 4.99 V, giving nonlinearity 0.84% of FSO.
A pressure transducer was calibrated with the following data.
| Pressure (bar) | 0 | 2 | 4 | 6 | 8 | 10 |
|---|---|---|---|---|---|---|
| Output (V) | 0.02 | 0.98 | 2.05 | 3.00 | 3.98 | 5.01 |
Fit the least-squares straight line, find the sensitivity and the zero offset, and calculate the maximum nonlinearity as a percentage of full-scale output (FSO).
Method: fit by least squares, then find the largest difference between the data and the line.
Least-squares fit
For points: , , , (units bar, V).
Here .
So the line is .
| x (bar) | y (V) | Fitted (V) | Deviation (V) |
|---|---|---|---|
| 0 | 0.02 | 0.0138 | +0.0062 |
| 2 | 0.98 | 1.0110 | -0.0310 |
| 4 | 2.05 | 2.0081 | +0.0419 |
| 6 | 3.00 | 3.0052 | -0.0052 |
| 8 | 3.98 | 4.0024 | -0.0224 |
| 10 | 5.01 | 4.9995 | +0.0105 |
Nonlinearity
Maximum deviation V at 4 bar.
Full-scale output V.
Answer: sensitivity (slope) V/bar, zero offset V, maximum nonlinearity of FSO.
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Zero drift and sensitivity drift · Instrumentation and Measurement, chapter 3
- Potentiometer transducer and loading error · Instrumentation and Measurement, chapter 5
- Second-order transducer frequency and step response · Instrumentation and Measurement, chapter 4
- Least squares fit for calibration data · Metrology, chapter 2
- Fit straight line: humidity and moisture · Probability and Statistics, chapter 5
- Quadratic least squares fit · Numerical Method, chapter 4
- Piezoelectric transducer sensitivity and output voltage · Instrumentation I, chapter 3