Zero drift and sensitivity drift
Short answer
Zero drift is a parallel shift of the calibration line; sensitivity drift is a change in slope. For the transducer, nominal output 30 mV plus 4 mV zero drift plus 1.2 mV from sensitivity drift gives total output 35.2 mV at 6 bar and 40°C.
(a) Explain sensitivity to disturbance, with zero drift and sensitivity drift (scale-factor drift) as the two forms.
(b) A pressure transducer has a sensitivity of 5 mV/bar at 20 C. The zero drift is 0.2 mV/C and the sensitivity drift is 0.01 mV/bar per C. Find the output when it reads 6 bar at 40 C.
(a) Sensitivity to disturbance
Environmental inputs such as ambient temperature, supply voltage, vibration or humidity can change the output of an instrument even when the measurand is constant. These are called disturbances (interfering and modifying inputs). Two effects are described.
- Zero drift (zero shift): the whole calibration line moves up or down by a constant amount, so the output at zero input changes. It is quoted per unit disturbance, e.g. mV/C.
- Sensitivity drift (scale-factor drift): the slope of the calibration line changes, so the error grows with input. Quoted as, e.g., (mV/bar) per C.
Output | / <- sensitivity drift (slope change)
| //
| ///_/ <- zero drift (parallel shift)
|_/_/
+---------------- Input
(b) Numerical
Temperature change C.
Nominal output mV.
Zero drift mV.
Sensitivity drift effect mV.
Answer: output mV, an error of mV (17%) due to temperature.
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Calibration line and nonlinearity of transducer · Instrumentation and Measurement, chapter 3
- Second-order transducer frequency and step response · Instrumentation and Measurement, chapter 4
- Potentiometer transducer and loading error · Instrumentation and Measurement, chapter 5
- Gauge resistance change and quarter-bridge output · Instrumentation and Measurement, chapter 6
- Models of zero, first and second order systems · Instrumentation and Measurement, chapter 4
- Full-bridge cantilever load cell output · Instrumentation and Measurement, chapter 6
- Piezoelectric transducer sensitivity and output voltage · Instrumentation I, chapter 3
- Tracking, disturbance rejection and pole sensitivity · Control Systems, chapter 3
- Two-wire 40 Ω line, input and output voltage · Electromagnetism, chapter 5
- Disturbance sensitivity and effect of gain · Control Systems, chapter 1