Static and dynamic circulating calorimetry; measurables

Short answer

Static calorimetry: insulated fluid mass m absorbs RF for time t; P = m c ΔT/t. Dynamic calorimetry: fluid flows at rate v; P = ρ v c (T_out − T_in). Key measurables/tools: power (bolometer/calorimeter), frequency (cavity wavemeter), VSWR/impedance (slotted line), S‑parameters (VNA), noise (noise figure meter).

Explain how microwave medium power is measured with the static and dynamic circulating calorimetry. List all the microwave measurables and their respective measuring tools.

A calorimeter measures microwave power by converting it fully into heat in a matched load and measuring the temperature rise. It is used for medium and high power (about 10 mW up to kW).

Static calorimetry

In a static calorimeter the load (a fixed mass of water, oil or a solid absorber) is thermally insulated and does not flow.

 RF in ===> [ matched load in insulated ] -- thermometer
            [ fixed mass m, sp. heat c  ]
  1. Microwave power is absorbed in the load for time t.
  2. The temperature rises by ΔT, read with a thermometer or thermocouple.
  3. Power absorbed:
P = m · c · ΔT / t     (watts)

Example: 0.2 kg of water (c = 4186 J/kg·°C) warms by 5 °C in 120 s → P = 0.2 × 4186 × 5 / 120 ≈ 34.9 W.

The method is simple but slow, and heat loss to the surroundings must be corrected.

Dynamic (circulating / flow) calorimetry

The fluid (usually water or oil) flows continuously through the load. In steady state all absorbed power is carried away by the fluid.

cold fluid in --> [ matched RF load ] --> hot fluid out
   T_in              ^ RF power           T_out
   flow rate v       |
  1. The fluid enters at T_in and leaves at T_out.
  2. With flow rate v (volume/s), density ρ and specific heat c:
P = ρ · v · c · (T_out − T_in)
For water: P (W) ≈ 4.187 × v (cm³/s) × ΔT (°C)
  1. The flow rate is fixed with a flowmeter; ΔT is measured by thermocouples or a temperature bridge.

Being continuous, it suits high power (kW) transmitters and has a faster reading than the static method.

Microwave measurables and tools

Quantity measuredMeasuring tool
Power (low)Bolometer/thermistor bridge, diode detector
Power (medium/high)Thermocouple, calorimeter wattmeter
FrequencyCavity wavemeter, frequency counter
Guide wavelengthSlotted line with probe
VSWR / reflection coefficientSlotted line + VSWR meter, reflectometer
ImpedanceSlotted line + Smith chart, network analyzer
Attenuation / insertion lossPower ratio or RF substitution method
S-parametersVector network analyzer
Q of cavitySwept frequency / transmission method
Noise figureNoise source + noise figure meter
SpectrumSpectrum analyzer