- 2081 Bhadra · 6+3 marks
Static and dynamic circulating calorimetry; measurables
Short answer
Static calorimetry: insulated fluid mass m absorbs RF for time t; P = m c ΔT/t. Dynamic calorimetry: fluid flows at rate v; P = ρ v c (T_out − T_in). Key measurables/tools: power (bolometer/calorimeter), frequency (cavity wavemeter), VSWR/impedance (slotted line), S‑parameters (VNA), noise (noise figure meter).
Explain how microwave medium power is measured with the static and dynamic circulating calorimetry. List all the microwave measurables and their respective measuring tools.
A calorimeter measures microwave power by converting it fully into heat in a matched load and measuring the temperature rise. It is used for medium and high power (about 10 mW up to kW).
Static calorimetry
In a static calorimeter the load (a fixed mass of water, oil or a solid absorber) is thermally insulated and does not flow.
RF in ===> [ matched load in insulated ] -- thermometer
[ fixed mass m, sp. heat c ]
- Microwave power is absorbed in the load for time t.
- The temperature rises by ΔT, read with a thermometer or thermocouple.
- Power absorbed:
P = m · c · ΔT / t (watts)
Example: 0.2 kg of water (c = 4186 J/kg·°C) warms by 5 °C in 120 s → P = 0.2 × 4186 × 5 / 120 ≈ 34.9 W.
The method is simple but slow, and heat loss to the surroundings must be corrected.
Dynamic (circulating / flow) calorimetry
The fluid (usually water or oil) flows continuously through the load. In steady state all absorbed power is carried away by the fluid.
cold fluid in --> [ matched RF load ] --> hot fluid out
T_in ^ RF power T_out
flow rate v |
- The fluid enters at T_in and leaves at T_out.
- With flow rate v (volume/s), density ρ and specific heat c:
P = ρ · v · c · (T_out − T_in)
For water: P (W) ≈ 4.187 × v (cm³/s) × ΔT (°C)
- The flow rate is fixed with a flowmeter; ΔT is measured by thermocouples or a temperature bridge.
Being continuous, it suits high power (kW) transmitters and has a faster reading than the static method.
Microwave measurables and tools
| Quantity measured | Measuring tool |
|---|---|
| Power (low) | Bolometer/thermistor bridge, diode detector |
| Power (medium/high) | Thermocouple, calorimeter wattmeter |
| Frequency | Cavity wavemeter, frequency counter |
| Guide wavelength | Slotted line with probe |
| VSWR / reflection coefficient | Slotted line + VSWR meter, reflectometer |
| Impedance | Slotted line + Smith chart, network analyzer |
| Attenuation / insertion loss | Power ratio or RF substitution method |
| S-parameters | Vector network analyzer |
| Q of cavity | Swept frequency / transmission method |
| Noise figure | Noise source + noise figure meter |
| Spectrum | Spectrum analyzer |
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Power measurement with circulating calorimeter · RF and Microwave Engineering, chapter 8 · asked 2×
- Microwave vs low frequency measurement; static calorimeter · RF and Microwave Engineering, chapter 8
- Power measurement with static calorimeter · RF and Microwave Engineering, chapter 8 · asked 2×
- Device to measure 7.5 mW microwave power · RF and Microwave Engineering, chapter 8
- Microwave vs traditional systems and need of S-parameters · RF and Microwave Engineering, chapter 1
- Static and dry calorimeters · RF and Microwave Engineering, chapter 8
- Static vs dynamic mathematical model · Simulation and Modeling, chapter 2 · asked 3×
- Static vs dynamic implementation of list · Data Structure and Algorithm, chapter 3 · asked 2×
- Static and dynamic hazards and their elimination · Digital Logic, chapter 3 · asked 2×
- Static versus dynamic binding trade-offs · Object Oriented Programming, chapter 7