- 2080 Baisakh · 6+2 marks
Calorimeter wattmeter and single bridge bolometer limits
Short answer
A calorimeter wattmeter uses a flowing liquid through a matched RF load; temperature rise is balanced against a DC‑heated reference, so RF power equals known DC power (or P = ρ v c ΔT). Single‑bridge bolometer limitations: ambient drift, ~10 mW max, manual balancing, mismatch error, DC–RF substitution error, and only average power.
Explain power measurement using calorimeter wattmeter. What are the limitations of using single bridge bolometer?
A calorimeter wattmeter measures high microwave power (watts to kilowatts) by absorbing it in a fluid load and comparing the heating with a known low-frequency or DC power.
Calorimeter wattmeter
Most practical calorimeter wattmeters are flow (circulating) type with a substitution balance:
+------------+ +------------+
oil/water->| RF load |------>| Input head |--+
stream | (absorbs | | (temp. | |
| RF power) | | sensor) | | bridge
+------------+ +------------+ +--> null
+------------+ +------------+ | meter
same -> | DC/LF load |------>| Ref. head |--+
stream | (heater) | | (temp. |
+------------+ | sensor) |
+------------+
Working principle:
- A liquid (water or oil) is pumped at a constant rate through the RF load, which is matched to the line and absorbs all the microwave power.
- The liquid heats up; its temperature rise is sensed by a temperature-sensitive element (thermistor or thermocouple) in the input head.
- A second, identical path has a reference head heated by a known DC or 60 Hz power.
- Both heads form arms of a bridge. The DC power is adjusted until the bridge balances, meaning both streams have the same temperature rise.
- Then RF power = known DC power, read on an ordinary wattmeter.
Without the reference, power can be found from the flow directly:
P = ρ · v · c · ΔT
Water: P ≈ 4.187 · v(cm³/s) · ΔT W
Features: wide range (10 mW to several kW), accuracy about ±1–5 %, broadband (load is well matched), but slow response and bulky (pump, plumbing).
Limitations of the single bridge bolometer
- Ambient temperature drift: the bolometer also responds to room temperature, so zero drifts; there is no compensating element.
- Low power only: limited to about 10 mW; higher power burns out the barretter or thermistor.
- Manual balancing: each reading needs the DC to be adjusted by hand, so it is slow and not direct reading.
- Mismatch error: bolometer resistance changes with power, so the mount match and reflected power change.
- DC–RF substitution error: RF and DC currents do not heat the element in exactly the same way (skin effect, current distribution).
- Average power only: it cannot show pulse peak power directly.
✓ Answer checked by Claude Opus 5.5 (high effort)
Related questions
- Power measurement with circulating calorimeter · RF and Microwave Engineering, chapter 8 · asked 2×
- Power measurement with static calorimeter · RF and Microwave Engineering, chapter 8 · asked 2×
- Double channel bolometer bridge method · RF and Microwave Engineering, chapter 8 · asked 2×
- VSWR meter and low power (bolometer) measurement · RF and Microwave Engineering, chapter 8 · asked 2×
- Microwave vs low frequency measurement; static calorimeter · RF and Microwave Engineering, chapter 8
- Power measurement tool for airport surveillance radar · RF and Microwave Engineering, chapter 8
- Single-phase electrodynamometer wattmeter · Instrumentation I, chapter 7
- Bridge for high-Q inductance measurement · Instrumentation I, chapter 2
- Bridge single-phase cycloconverter conduction sequence · Power Electronics, chapter 6 · asked 3×
- Single-phase diode bridge with constant load current · Power Electronics, chapter 2